| ¡¡ |

|
|
|
|
>>
Cardio Products--ECG Knowledge
|
|
►Electrocardiogram |
►Electrocardio-graph description |
|
►The
electrocardiogram |
►Electrocardio-graph DEFINITION |
|
►The
Art of The ECG / EKG Machine |
►Electrocardio-graph graph paper |
|
►A brief
history of electrocardiography |
►Interpreting Electrocardio-graph |
|
►History and
Term Electrocardiogram |
| ¡¡ |
| ¡¡ |
|
Electrocardio-graph DEFINITION |
Electrocardiography (ECG or EKG) is a transthoracic
interpretation of the electrical activity of the heart over time
captured and externally recorded by skin electrodes.[1] It is a
noninvasive recording produced by an electrocardiographic
device. The etymology of the word is derived from the Greek
electro, because it is related to electrical activity, cardio,
Greek for heart, and graph, a Greek root meaning "to write". In
English speaking countries, medical professionals often write
EKG (the German abbreviation) in order to avoid confusion with
EEG.
The ECG works mostly by detecting and amplifying the tiny
electrical changes on the skin that are caused when the heart
muscle "depolarises" during each heart beat. At rest, each heart
muscle cell has a charge across its outer wall, or cell
membrane. Reducing this charge towards zero is called de-polarisation,
which activates the mechanisms in the cell that cause it to
contract. During each heartbeat a healthy heart will have an
orderly progression of a wave of depolarisation that is
triggered by the cells in the sinoatrial node, spreads out
through the atrium, passes through "intrinsic conduction
pathways" and then spreads all over the ventricles. This is
detected as tiny rises and falls in the voltage between two
electrodes placed either side of the heart which is displayed as
a wavy line either on a screen or on paper. This display
indicates the overall rhythm of the heart and weaknesses in
different parts of the heart muscle.
Usually more than 2 electrodes are used and they can be combined
into a number of pairs. (For example: Left arm (LA),right arm
(RA) and left leg (LL) electrodes form the pairs: LA+RA, LA+LL,
RA+LL) The output from each pair is known as a lead. Each lead
is said to look at the heart from a different angle. Different
types of ECGs can be referred to by the number of leads that are
recorded, for example 3-lead, 5-lead or 12-lead ECGs (sometimes
simply "a 12-lead"). A 12-lead ECG is one in which 12 different
electrical signals are recorded at approximately the same time
and will often be used as a one-off recording of an ECG,
typically printed out as a paper copy. 3- and 5-lead ECGs tend
to be monitored continuously and viewed only on the screen of an
appropriate monitoring device, for example during an operation
or whilst being transported in an ambulance. There may, or may
not be any permanent record of a 3- or 5-lead ECG depending on
the equipment used.
It is the best way to measure and diagnose abnormal rhythms of
the heart,[2] particularly abnormal rhythms caused by damage to
the conductive tissue that carries electrical signals, or
abnormal rhythms caused by electrolyte imbalances.[3] In a
myocardial infarction (MI), the ECG can identify if the heart
muscle has been damaged in specific areas, though not all areas
of the heart are covered.[4] The ECG cannot reliably measure the
pumping ability of the heart, for which ultrasound-based
(echocardiography) or nuclear medicine tests are used. It is
possible to be in cardiac arrest with a normal ECG signal (a
condition known as pulseless electrical activity). |
| Article Source: |
| http://en.wikipedia.org/wiki/Electrocardiography |
|
Back To Top |
| ¡¡ |
|
Electrocardio-graph graph paper |
One second of ECG graph paperThe output of an ECG recorder
is a graph (or sometimes several graphs, representing each of
the leads) with time represented on the x-axis and voltage
represented on the y-axis. A dedicated ECG machine would usually
print onto graph paper which has a background pattern of 1mm
squares (often in red or green), with bold divisions every 5mm
in both vertical and horizontal directions. It is possible to
change the output of most ECG devices but it is standard to
represent each mV on the y axis as 1 cm and each second as 25mm
on the x-axis (that is a paper speed of 25mm/s). Faster paper
speeds can be used - for example to resolve finer detail in the
ECG. At a paper speed of 25 mm/s, one small block of ECG paper
translates into 40 ms. Five small blocks make up one large
block, which translates into 200 ms. Hence, there are five large
blocks per second. A calibration signal may be included with a
record. A standard signal of 1 mV must move the stylus
vertically 1 cm, that is two large squares on ECG paper.
Layout
By definition a 12-lead ECG will show a short segment of the
recording of each of the 12-leads. This is often arranged in a
grid of 4 columns by three rows, the first columns being the
limb leads (I,II and III), the second column the augmented limb
leads (aVR, aVL and aVF) and the last two columns being the
chest leads (V1-V6). It is usually possible to change this
layout so it is vital to check the labels to see which lead is
represented. Each column will usually record the same moment in
time for the three leads and then the recording will switch to
the next column which will record the heart beats after that
point. It is possible for the heart rhythm to change between the
columns of leads. Each of these segments is short, perhaps 1-3
heart beats only, depending on the heart rate and it can be
difficult to analyse any heart rhythm that shows changes between
heart beats. To help with the analysis it is common to print one
or two "rhythm strips" as well. This will usually be lead II
(which shows the electrical signal from the atrium, the P-wave,
well) and shows the rhythm for the whole time the ECG was
recorded (usually 5¨C6 seconds). The term "rhythm strip" may also
refer to the whole printout from a continuous monitoring system
which may show only one lead and is either initiated by a
clinician or in response to an alarm or event.
Leads
The term "lead" in electrocardiography causes much confusion
because it is used to refer to two different things. In
accordance with common parlance the word lead may be used to
refer to the electrical cable attaching the electrodes to the
ECG recorder. As such it may be acceptable to refer to the "left
arm lead" as the electrode (and its cable) that should be
attached at or near the left arm. There are usually ten of these
electrodes in a standard "12-lead" ECG.
Alternatively (and some would say properly, in the context of
electrocardiography) the word lead may refer to the tracing of
the voltage difference between two of the electrodes and is what
is actually produced by the ECG recorder. Each will have a
specific name. For example "Lead I" (lead one) is the voltage
between the right arm electrode and the left arm electrode,
whereas "Lead II" (lead two) is the voltage between the right
limb and the feet. (This rapidly becomes more complex as one of
the "electrodes" may in fact be a composite of the electrical
signal from a combination of the other electrodes. (See later.)
Twelve of this type of lead form a "12-lead" ECG
To cause additional confusion the term "limb leads" usually
refers to the tracings from leads I, II and III rather than the
electrodes attached to the limbs.
Small Text===Placement of electrodes=== Ten electrodes are used
for a 12-lead ECG. The electrodes usually consist of a
conducting gel, embedded in the middle of a self-adhesive pad
onto which cables clip. Sometimes the gel also forms the
adhesive.[12] They are labeled and placed on the patient's body
as follows:
| Electrode label (in the USA) |
Electrode placement |
| RA |
On the right arm, avoiding bony prominences. |
| LA |
In the same location that RA was placed, but on the
left arm this time. |
| RL |
On the right leg, avoiding bony prominences. |
| LL |
In the same location that RL was placed, but on the
left leg this time. |
| V1 |
In the fourth intercostal space (between ribs
4 & 5) just to the right of the sternum
(breastbone). |
| V2 |
In the fourth intercostal space (between ribs
4 & 5) just to the left of the sternum. |
| V3 |
Between leads V2 and V4. |
| V4 |
In the fifth intercostal space (between ribs 5 & 6)
in the mid-clavicular line (the imaginary line that
extends down from the midpoint of the clavicle
(collarbone)). |
| V5 |
Horizontally even with V4, but in the
anterior axillary line. (The anterior axillary line is
the imaginary line that runs down from the point midway
between the middle of the clavicle and the lateral end
of the clavicle; the lateral end of the collarbone is
the end closer to the arm.) |
| V6 |
Horizontally even with V4 and V5
in the midaxillary line. (The midaxillary line is the
imaginary line that extends down from the middle of the
patient's armpit.) |
|
| Article Source: |
| http://en.wikipedia.org/wiki/Electrocardiography |
|
Back To Top |
| ¡¡ |
|
A brief history of
electrocardiography |
1600
William Gilbert William Gilbert, Physician to Queen Elizabeth I,
President of the College of Physicians (before its Royal
Charter), and creator of the 'magnetic philosophy' introduces
the term 'electrica' for objects (insulators) that hold static
electricity. He derived the word from the Greek for amber (electra).
It was known from ancient times that amber when rubbed could
lift light materials. Gilbert added other examples such as
sulphur and was describing what would later be known as 'static
electricity' to distinguish it from the more noble magnetic
force which he saw as part of a philosophy to destroy forever
the prevailing Aristotlean view of matter. Gilbert W. De Magnete,
magneticisique corporibus, et de magno magnete tellure. [On the
Magnet, Magnetic Bodies, and the Great Magnet of the Earth] 1600
1646
Sir Thomas Browne, Physician, whilst writing to dispel popular
ignorance in many matters, is the first to use the word
'electricity'. Browne calls the attractive force "Electricity,
that is, a power to attract strawes or light bodies, and convert
the needle freely placed". (He is also the first to use the word
'computer' - referring to people who compute calendars.) Browne,
Sir Thomas. Pseudodoxia Epidemica: Or, enquiries Into Very Many
Received Tenents, and Commonly Presumed Truths. 1646: Bk II, Ch.
1. London
1660
Otto Von Guericke builds the first static electricity generator.
1662
Descarte's reflex ©BIU The work of Rene Descartes, French
Philosopher, is published (after his death) and explains human
movement in terms of the complex mechanical interaction of
threads, pores, passages and 'animal spirits'. He had worked on
his ideas in the 1630s but had abandoned publication because of
the persecution of other radical thinkers such as Galileo.
William Harvey had developed similar ideas but they were never
published. Descartes R. De Homine (Treatise of Man); 1662:
Moyardum & leffen, Leiden.
1664
Jan Swammerdam, a Dutchman, disproves Descartes' mechanistic
theory of animal motion by removing the heart of a living frog
and showing that it was still able to swim. On removing the
brain all movement stopped (which would be in keeping with
Descarte's theory) but then, when the frog was dissected and a
severed nerve end stimulated with a scalpel the muscles
twitched. This proved that movement of a muscle could occur
without any connection to the brain and therefore the
transmission of 'animal spirits' was not necessary.
Swammerdam's ideas were not widely known and his work was not
published until after his death. However, he wrote many letters
and his friend, Nicolaus Steno, did attack the Cartesian ideas
in a lecture in Paris in 1665. Boerhaave published Swammerdam's
'Book of Nature' in the 1730s which was translated into English
in 1758.
1668
electrical stimulation? ©BIU Swammerdam refines his experiments
on muscle contraction and nerve conduction and demonstrated some
to notable figures such as the Grand-Duke Cosimo of Tuscany who
was visiting Swammerdam's father's house on the Oude Schans in
Amsterdam. One experiment suspended the muscle on a brass hook
inside a glass tube with a water droplet to detect movement and
'irritated' the nerve with a silver wire. This produced movement
of the muscle and it may have been due to the induction of a
small electrical charge - although Swammerdam would have been
unaware of this.
In the diagram opposite - a) glass tube, b) muscle, c) sliver
wire, d) brass wire, e) drop of water, f) investigator's hand.
1729
Stephen Gray, English scientist, distinguishes between
conductors and insulators of electricity. He demonstrates the
transfer of static electrical charge to a cork ball across 150
metres of wet hemp thread. Later he found that the transfer
could be achieved over greater distances by using brass wire.
1745
Leyden Jar Dutch physicist Pieter van Musschenbroek discovers
that a partly filled jar with a nail projecting from a cork in
its neck can store an electrical charge. The jar is named the 'Leyden
Jar' after the place of its discovery. Ewald Georg von Kliest of
Pomerania invented the same device independently.
Using a Leyden jar in 1746, Jean-Antoine Nollet, French
physicist and tutor to the Royal family of France sends an
electrical current through 180 Royal Guards during a
demonstration to King Louis XV.
1769
Edward Bancroft, an American Scientist, suggests that the
'shock' from the Torpedo Fish is electrical rather than
mechanical in nature. He showed that the properties of the shock
were similar to those from a Leyden jar in that it could be
conducted or insulated with appropriate materials. The Torpedo
fish and other species were widely known to deliver shocks and
were often used in this way for therapeutic reasons. However,
electrical theory at the time dictated that electricity would
always flow through conductors and diffuse away from areas of
high charge to low charge. Since living tissues were known to be
conductors it was impossible to imagine how an imbalance of
charge could exist within an animal and therefore animals could
not use electricity for nerve conduction - or to deliver shocks.
Furthermore, 'water and electricity do not mix' so the idea of
an 'electric fish' was generally not accepted. Bancroft, E. An
essay on the natural history of Guiana, London:T. Becket and P.
A. de Hondt, 1769.
1773
John Walsh John Walsh, fellow of the Royal Society and Member of
Parliament, obtains a visible spark from an electric eel
Electrophorus electricus. The eel was out of water as it was not
possible to produce the spark otherwise. He used thin strips of
tin foil and demonstrated his technique to many colleagues and
visitors at his house in London. Unfortunately he never
published his eel experiment though he did win the Copley medal
in 1774 and 1783 for his work. The observations of Walsh, and
Bancroft before him, added to the argument that some form of
animal electricity existed. Walsh, J. On the electric property
of torpedo: in a letter to Ben. Franklin. Phil. Trans. Royal
Soc. 1773;63:478-489
1774
The Rev. Mr Sowdon and Mr Hawes, apothecary, report on the
surprising effects of electricity in a case report of recovery
from sudden death published in the annual report of the newly
founded Humane Society now the Royal Humane Society. The Society
had developed from 'The Institution for Affording immediate
relief to persons apparently dead from drowning'. It was
"instituted in the year 1774, to protect the industrious from
the fatal consequences of unforseen accidents; the young and
inexperienced from being sacrificed to their recreations; and
the unhappy victims of desponding melancholy and deliberate
suicide; from the miserable consequences of self-destruction."
A Mr Squires, of Wardour Street, Soho lived opposite the house
from which a three year old girl, Catherine Sophia Greenhill had
fallen from the first storey window on 16th July 1774. After the
attending apothecary had declared that nothing could be done for
the child Mr Squires, "with the consent of the parents very
humanely tried the effects of electricity. At least twenty
minutes had elapsed before he could apply the shock, which he
gave to various parts of the body without any apparent success;
but at length, upon transmitting a few shocks through the
thorax, he perceived a small pulsation: soon after the child
began to sigh, and to breathe, though with great difficulty. In
about ten minutes she vomited: a kind of stupor, occaisioned by
the depression of the cranium, remained for some days, but
proper means being used, the child was restored to perfect
health and spirits in about a week.
"Mr. Squires gave this astonishing case of recovery to the above
gentlemen, from no other motive than a desire of promoting the
good of mankind; and hopes for the future that no person will be
given up for dead, till various means have been used for their
recovery."
Since it is clear she sustained a head injury the electricity
probably stimulated the child out of deep coma rather than
providing cardiac defibrillation (see also 1788, Charles Kite).
Annual Report 1774: Humane Society, London. pp 31-32
1775
Abildgaard shows that hens can be made lifeless with electrical
impulses and he could restore a pulse with electrical shocks
across the chest. "With a shock to the head, the animal was
rendered lifeless, and arose with a second shock to the chest;
however, after the experiment was repeated rather often, the hen
was completely stunned, walked with some difficulty, and did not
eat for a day and night; then later it was very well and even
laid an egg." Abildgaard, Peter Christian. Tentamina electrica
in animalibus. Inst Soc Med Havn. 1775; 2:157-61.
1786
Luigi Galvani Italian Anatomist Luigi Galvani notes that a
dissected frog's leg twitches when touched with a metal scalpel.
He had been studying the effects of electricity on animal
tissues that summer.
On 20th September 1786 he wrote "I had dissected and prepared a
frog in the usual way and while I was attending to something
else I laid it on a table on which stood an electrical machine
at some distance from its conductor and separated from it by a
considerable space. Now when one of the persons present touched
accidentally and lightly the inner crural nerves of the frog
with the point of a scalpel, all the muscles of the legs seemed
to contract again and again as if they were affected by powerful
cramps."
He later showed that direct contact with the electrical
generator or the ground through an electrical conductor would
lead to a muscle contraction. Galvani also used brass hooks that
attached to the frog's spinal cord and were suspended from an
iron railing in a part of his garden. He noticed that the frogs'
legs twitched during lightening storms and also when the weather
was fine. He interperated these results in terms of "animal
electricity" or the preservation in the animal of "nerveo-electrical
fluid" similar to that of an electric eel. He later also showed
that electrical stimulation of a frog's heart leads to cardiac
muscular contraction. Galvani. De viribus Electritatis in motu
musculari Commentarius. 1791
Galvani's name is given to the 'galvanometer' which is an
instrument for measuring (and recording) electricity - this is
essentially what an ECG is; a sensitive galvanometer.
1788
Charles Kite wins the Silver Medal of the Humane Society
(awarded at the first Prize Medal ceremony of the Society
co-judged with the Medical Society of London) with an essay on
the use of electricity in the diagnosis and resuscitation of
persons apparently dead. This essay is often cited as the first
record of cardiac defibrillation but the use of electricity
suggested by Mr Kite is much different. For example, on
describing a case of drowning from 1785 where resuscitation had
been attempted with artificial respiration, warmth, tobacco,
"volatiles thrown into the stomach, frictions, and various
lesser stimuli" for nearly an hour, he then recalls the use of
electricity. "Electricity was then applied, and shocks sent
through in every possible direction; the muscles through which
the fluid [electricity] passed were thrown into strong
contractions." He concluded that electricity was a valuable tool
that could determine whether or not a person, apparently dead,
could be successfully resuscitated. Annual Report 1788: Humane
Society, London. pp 225-244. Kite C. An Essay on the Recovery of
the Apparently Dead. 1788: C. Dilly, London.
1792
Alessandro Volta Alessandro Volta, Italian Scientist and
inventor, attempts to disprove Galvani's theory of "animal
electricity'" by showing that the electrical current is
generated by the combination of two dissimilar metals. His
assertion was that the electrical current came from the metals
and not the animal tissues. (We now know that both Galvani and
Volta were right.) To prove his theory he develops the voltaic
pile in 1800 (a column of alternating metal discs - zinc with
copper or silver - separated by paperboard soaked in saline)
which can deliver a substantial and steady current of
electricity. Enthusiasm in the use of electricity leads to
further attempts at reanimation of the dead with experiments on
recently hanged criminals. Giovani Aldini (the nephew of Galvani)
conducts an experiment at the Royal College of Surgeons in
London in 1803. The executed criminal had lain in a temperature
of 30 F for one hour and was transported to the College. "On
applying the conductors to the ear and to the rectum, such
violent muscular contractions were executed, as almost to give
the appearance of the reanimation". Aldini, J. Essai: ThöVique et
expöémental sur le Galvanisme, Paris (1804), Giovani Aldini.
General Views on the Application of Galvanism to Medical
Purposes Principally in cases of suspended Animation (London: J.
Callow, Princes Street and Burgess and Hill, Great Windmill
Street, 1819). Mary Shelly's Frankenstein was published in 1818.
Louis Figuier, Les merveilles de la Science (Paris, 1867), p.653
1800 to 1895 The design of sensitive instruments that could
detect the small electrical currents in the heart.
1819
While demonstrating to students the heating of a platinum wire
with electricity from a voltaic pile at the University of
Copenhagen, Danish physicist Hans Christian Oersted notices that
a nearby magnetized compass needle moves each time the
electrical current is turned on. He discovers electromagnetism
which is given a theoretical basis (with remarkable speed) by
AndríŸarie Ampؘ.
1820
Johann (Johan) Schweigger of Nuremberg increases the movement of
magnetized needles in electromagnetic fields. He found that by
wrapping the electric wire into a coil of 100 turns the effect
on the needle was multiplied. He proposed that a magnetic field
revolved around a wire carrying a current which was later proven
by Michael Faraday. Schweigger had invented the first
galvanometer and announced his discovery at the University of
Halle on 16th September 1820.
1825
Leopoldo Nobili, Professor of Physics at Florence, develops an
'astatic galvanometer'. Using two identical magnetic needles of
opposite polarity, either fixed together with a figure of eight
arrangment of wire loops (in earlier versions), or one moveable
needle with a wire loop and one with a scale (in later
versions), the effects of the earth's magnetic field could be
compensated for. In 1827, using this instrument, he managed to
detect the flow of current in the body of a frog from muscles to
spinal cord. He detected the electricity running along saline
moistened cotton thread joining the dissected frog's legs in one
jar to its body in another jar. Nobili was working to support
the theory of animal electricity and this conduction,
transmitted without wires, he felt demonstrated animal
electricity.
1838
Carlo Matteucci Carlo Matteucci, Professor of Physics at the
University of Pisa, and student of Nobili, shows that an
electric current accompanies each heart beat. He used a
preparation known as a 'rheoscopic frog' in which the cut nerve
of a frog's leg was used as the electical sensor and twitching
of the muscle was used as the visual sign of electrical
activity. He also used Nobili's astatic galvanometer for the
study of electricity in muscles typically inserting one
galvanometer wire in the open end of the dissected muscle and
the other on the surface of the muscle. He went on to try and
demonstrate conduction in nerve but was unable to do so (since
his galvanometers were not sensitive enough). Matteucci C. Sur
un phenomene physiologique produit par les muscles en
contraction. Ann Chim Phys 1842;6:339-341
1840
Dr Golding Bird, a Physician, accomplished chemist and member of
the London Electrical Society, opens an electrical therapy room
at Guy's Hospital, London treating a large range of diseases.
Although the application of electricity was popular it was not
considered a subject worthy of serious investigation. Because of
Bird's reputation as a researcher electrical therapy achieved
popularity amongst London Physicians including his mentor Dr
Thomas Addison. Bird G. Lectures on Electricity and Galvanism,
in their physiological and therapeutical relations, delivered at
the Royal College of Physicians, in March, 1847 (Wilson &
Ogilvy, London, 1847)
1843
Emil Du bois-Reymond German physiologist Emil Du bois-Reymond
describes an "action potential" accompanying each muscular
contraction. He detected the small voltage potential present in
resting muscle and noted that this diminished with contraction
of the muscle. To accomplish this he had developed one of the
most sensitive galvanometers of his time. His device had a wire
coil with over 24,000 turns - 5 km of wire. Du Bios Reymond
devised a notation for his galvanometer which he called the
'disturbance curve'. "o" was the stable equilibrium point of the
astatic galvanometer needle and p, q, r and s (and also k and h)
were other points in its deflection. Du Bois-Reymond, E.
Untersuchungen uber thierische Elektricitat. Reimer, Berlin:
1848.
1849
H. Bence Jones, Physician at St George's Hospital, reports on
the disappointing results of electrical therapy on his patients.
He concluded that only four of his 23 cases seemed to improve.
H. Bence Jones. Remedial Action of Electricity. Lond J Med Feb
1849; s2-1: 125 - 129; doi:10.1136/bmj.s2-1.2.125
1850
Bizarre unregulated actions of the ventricles (later called
ventricular fibrillation) is described by Hoffa during
experiments with strong electrical currents across the hearts of
dogs and cats. He demonstrated that a single electrical pulse
can induce fibrillation. Hoffa M, Ludwig C. 1850. Einige neue
versuche uber herzbewegung. Zeitschrift Rationelle Medizin, 9:
107-144
1856
Rudolph von Koelliker and Heinrich Muller confirm that an
electrical current accompanies each heart beat by applying a
galvanometer to the base and apex of an exposed ventricle. They
also applied a nerve-muscle preparation, similar to Matteucci's,
to the ventricle and observed that a twitch of the muscle
occured just prior to ventricular systole and also a much
smaller twitch after systole. These twitches would later be
recognised as caused by the electrical currents of the QRS and T
waves. von Koelliker A, Muller H. Nachweis der negativen
Schwankung des Muskelstroms am naturlich sich kontrahierenden
Herzen. Verhandlungen der Physikalisch-Medizinischen
Gesellschaft in Wurzberg. 1856;6:528-33.
1858
William Thompson (Lord Kelvin), Professor of Natural Philosophy
at Glasgow University, invents the 'mirror galvanometer' for the
reception of transatlantic telegraph transmissions. A small,
freely rotating mirror, with magents stuck to its back is
suspended in a fine copper coil and a reflected spot of light
from this mirror 'amplifies' the small movements when electrical
current is present. The whole apparatus was suspended in an air
chamber and the pressure inside could be adjusted to vary the
damping seen on the signals. This galvanometer was sensitive
enough for transatlantic telegraphy.
1867
Thompson improves telegraph transmissions with the 'Siphon
Recorder'. Before d'Arsonval (1880), Thompson uses a fine coil
suspended in a strong magnetic magnetic field. Attached to the
coil but isolated from it by ebonite (an insulator) was a siphon
of ink. The siphon was charged with high voltage so that the ink
was sprayed onto the paper that moved over an earthed metal
surface. The siphon recorder could therefore not only detect
currents it could also record them onto paper.
1869-70
Alexander Muirhead, an electrical engineer and pioneer of
telegraphy, may have a recorded a human electrocardiogram at St
Bartholomew's Hospital, London but this is disputed. If he had
he is thought to have used a Thompson Siphon Recorder. Elizabeth
Muirhead, his wife, wrote a book of his life and claimed that he
refrained from publishing his own work for fear of misleading
others. Elizabeth Muirhead. Alexander Muirhead 1848 - 1920.
Oxford, Blackwell: privately printed 1926.
1872
French physicist Gabriel Lippmann invents a capillary
electrometer. It is a thin glass tube with a column of mercury
beneath sulphuric acid. The mercury meniscus moves with varying
electrical potential and is observed through a microscope.
1872
Mr Green, a surgeon, publishes a paper on the resuscitation of a
series of patients who had suffered cardiac and / or respiratory
arrest during anaesthesia with chloroform. He uses a galvanic
pile (battery) of 200 cells generating 300 Volts which he
applied to the patient as follows "One pole should be applied to
the neck and the other to the lower rib on the left side." Green
T. On death from chloroform: its prevention by galvanism. Br Med
J 1872 1: 551-3. Although this has been reported as an example
of cardiorespiratory resuscitation it is unclear what the exact
mechanism seems to be. It is unlikely to be electric
cardioversion or external pacing. It seems to be another example
of electrophrenic stimulation (See also Duchenne 1872).
1872
An 'electric' smile. Guillaume Benjamin Amand Duchenne de
Boulogne, pioneering neurophysiologist, describes the
resuscitation of a drowned girl with electricity in the third
edition of his textbook on the medical uses of electricity. This
episode has sometimes been described as the first 'artificial
pacemaker' but he used an electrical current to induce
electrophrenic rather than myocardial stimulation. Duchenne GB.
De l'electrisation localisee et de son application a la
pathologie et la therapeutique par courants induits at par
courants galvaniques interrompus et continus. [Localised
electricity and its application to pathology and therapy by
means of induced and galvanic currents, interrupted and
continuous] 3ed. Paris. JB Bailliere et fils; 1872
1875
Richard Caton, a Liverpool Physician, presents to the British
Medical Association in July 1875 in Edinburgh. Using a Thompson
'mirror galvanometer' in animals he shows it was possible to
detect 'feeble currents of varying direction ... when the
electrodes are placed on two points of the external surface, or
one electrode on the grey matter and one on the surface of the
skull'. This is the first report of the EEG (or
electroencephalogram). Caton was exploring another Physician's
hypothesis, John Hughlings Jackson, who suggested in 1873 that
epilepsy was due to excessive electrical activity in the grey
matter of the brain. Caton R: The electric currents of the
brain. Br Med J 1875; 2(765):278, Mumenthaler, Mattle Eds.
Neurology. 4th Edition. Stuttgart, Thieme: 2004.
1876
Marey uses the electrometer to record the electrical activity of
an exposed frog's heart. Marey EJ. Des variations electriques
des muscles et du couer en particulier etudies au moyen de
l'electrometre de M Lippman. Compres Rendus Hebdomadaires des
Seances de l'Acadamie des sciences 1876;82:975-977
1878
British physiologists John Burden Sanderson and Frederick Page
record the heart's electrical current with a capillary
electrometer and shows it consists of two phases (later called
QRS and T). Burdon Sanderson J. Experimental results relating to
the rhythmical and excitatory motions of the ventricle of the
frog. Proc R Soc Lond 1878;27:410-414
1880
French physicist Ars¼¥ d'Arsonval in association with Marcel
Deprez, improves the galvanometer. Instead of a magnetized
needle moving when electrical current flows through a
surrounding wire coil the Deprez-d'Arsonval galvanometer has a
fixed magnet and moveable coil. If a pointer is attached to the
coil it can move over a suitably calibrated scale. The
d'Arsonval galvanometer is the basis for most modern
galvanometers. Comptes rendus de l'AcadôŽe des sciences, 1882,
94: 1347-1350
1884
John Burden Sanderson and Frederick Page publish some of their
recordings. Burdon Sanderson J, Page FJM. On the electrical
phenomena of the excitatory process in the heart of the
tortoise, as investigated photographically. J Physiol (London)
1884;4:327-338
1887
British physiologist Augustus D. Waller of St Mary's Medical
School, London publishes the first human electrocardiogram. It
is recorded with a capilliary electrometer from Thomas Goswell,
a technician in the laboratory. Waller AD. A demonstration on
man of electromotive changes accompanying the heart's beat. J
Physiol (London) 1887;8:229-234
1889
Dutch physiologist Willem Einthoven sees Waller demonstrate his
technique at the First International Congress of Physiologists
in Bale. Waller often demonstrated by using his dog "Jimmy" who
would patiently stand with paws in glass jars of saline.
1889
Professor John McWilliam, of Aberdeen University, describes
ventricular fibrillation as "unexpected, and irretrievable
cardiac failure [which] may ... present itself in the form of an
abrupt onset of fibrillar contraction ... The cardiac pump is
thrown out of gear, and the last of its vital energy is
dissipated in a violent and prolonged turmoil of fruitless
activity in the ventricular walls." He also describes the
electrical stimulation of the heart in cases of "fatal syncope"
in man. "A single induction shock readily causes a beat in an
inhibited heart, and a regular series of induction shocks (for
example, sixty or seventy per minute) gives a regular series of
heartbeats at the same rate." McWilliam JA. Cardiac Failure and
Sudden Death. Br Med J 1889;1:6-8. McWilliam JA. Electrical
stimulation of the heart in man. Br Med J 1889;1:348?50.
1890
GJ Burch of Oxford devises an arithmetical correction for the
observed (sluggish) fluctuations of the electrometer. This
allows the true waveform to be seen but only after tedious
calculations. Burch GJ. On a method of determining the value of
rapid variations of a difference potential by means of a
capillary electrometer. Proc R Soc Lond (Biol) 1890;48:89-93
1891
British physiologists William Bayliss and Edward Starling of
University College London improve the capillary electrometer.
They connect the terminals to the right hand and to the skin
over the apex beat and show a "triphasic variation accompanying
(or rather preceding) each beat of the heart". These deflections
are later called P, QRS and T. Bayliss WM, Starling EH. On the
electrical variations of the heart in man. Proc Phys Soc (14th
November) in J Physiol (London) 1891;13:lviii-lix and also On
the electromotive phenomena of the mammalian heart. Proc R Soc
Lond 1892;50:211-214 They also demonstrate a delay of about 0.13
seconds between atrial stimulation and ventricular
depolarisation (later called PR interval). On the electromotive
phenomena of the mammalian heart. Proc Phys Soc (21st March) in
J Physiol (London) 1891;12:xx-xxi
1893
Willem Einthoven introduces the term 'electrocardiogram' at a
meeting of the Dutch Medical Association. (Later he claims that
Waller was first to use the term). Einthoven W: Nieuwe methoden
voor clinisch onderzoek [New methods for clinical
investigation]. Ned T Geneesk 29 II: 263-286, 1893
1895 to date The first accurate recording of the
electrocardiogram and its development as a clinical tool.
1895
Einthoven, using an improved electrometer and a correction
formula developed independently of Burch, distinguishes five
deflections which he names P, Q, R, S and T. Einthoven W. Ueber
die Form des menschlichen Electrocardiogramms. Arch f d Ges
Physiol 1895;60:101-123
--------------------------------------------------------------------------------
Why PQRST and not ABCDE? The four deflections prior to the
correction formula were labelled ABCD and the 5 derived
deflections were labelled PQRST. The choice of P is a
mathematical convention dating from Descartes (as used also by
Du Bois-Reymond in his galvanometer's 'disturbance curve' 50
years previously) by using letters from the second half of the
alphabet. N has other meanings in mathematics and O is used for
the origin of the Cartesian coordinates. In fact Einthoven used
O ..... X to mark the timeline on his diagrams. P is simply the
next letter. A lot of work had been undertaken to reveal the
true electrical waveform of the ECG by eliminating the damping
effect of the moving parts in the amplifiers and using
correction formulae. If you look at the diagram in Einthoven's
1895 paper you will see how close it is to the string
galvanometer recordings and the electrocardiograms we see today.
The image of the PQRST diagram may have been striking enough to
have been adopted by the researchers as a true representation of
the underlying form. It would have then been logical to continue
the same naming convention when the more advanced string
galvanometer started creating electrocardiograms a few years
later. (For more on Descartes see Henson JR. Descartes and the
ECG lettering series. J Hist Med Allied Sci. April 1971;181?186)
--------------------------------------------------------------------------------
1897
Clement Ader, a French electrical engineer, reports his
amplification system for detecting Morse code signals
transmitted along undersea telegraph lines. It was never
intended to be used as a galvanometer. Einthoven later quoted
Ader's work but seems to have developed his own amplification
device independently. Ader C. Sur un nouvel appareil
enregistreur pour cables sous-marins. C R Acad Sci (Paris)
1897;124:1440-1442
1899
Karel Wenkebach Karel Frederik Wenckebach publishes a paper "On
the analysis of irregular pulses" describing impairment of AV
conduction leading to progressive lengthening and blockage of AV
conduction in frogs. This will later be called Wenckebach block
(Mobitz type I) or Wenckebach phenomenon.
1899
Jean-Louis Prevost, Professor of Biochemistry, and Frederic
Batelli, Professor of Physiology, both of Geneva discover that
large electrical voltages applied across an animal's heart can
stop ventricular fibrillation. Prevost JL, Batelli F. Sur
quelques effets des descharges electriques sur le coeur des
mammiferes. [On the effects of electric shocks on the hearts of
mammals.] Acad. Sci. Paris, FR.: 1899; 129:1267-1268.. They also
report that ventricular fibrillation can be induced by small
voltages (40 V). Prevost JL, Batelli F. La Mort Par Les
Descharges Electriques. [Death by electrical disharges]. Journ.
de Physiol. 1899;1:1085-99.
top
1901
Einthoven invents a new galvanometer for producing
electrocardiograms using a fine quartz string coated in silver
based on ideas by Deprez and d'Arsonval (who used a wire coil).
His "string galvanometer" weighs 600 pounds. Einthoven
acknowledged the similar system by Ader but later (1909)
calculated that his galvanometer was in fact many thousands of
times more sensitive. Einthoven W. Un nouveau galvanometre. Arch
Neerl Sc Ex Nat 1901;6:625-633
1902
Einthoven publishes the first electrocardiogram recorded on a
string galvanometer. Einthoven W. Galvanometrische registratie
van het menschilijk electrocardiogram. In: Herinneringsbundel
Professor S. S. Rosenstein. Leiden: Eduard Ijdo, 1902:101-107
1903
Einthoven discusses commercial production of a string
galvanometer with Max Edelmann of Munich and Horace Darwin of
Cambridge Scientific Instruments Company of London.
1905
Einthoven starts transmitting electrocardiograms from the
hospital to his laboratory 1.5 km away via telephone cables. On
March 22nd the first 'telecardiogram' is recorded from a healthy
and vigorous man and the tall R waves are attributed to his
cycling from laboratory to hospital for the recording.
1905
John Hay of Liverpool, publishes pressure recordings from a 65
year old man showing heart block in which AV conduction did not
seem to be impaired since the a-c intervals on the jugular
venous waves was unchanged in the conducted beats. This is the
first demonstration of what we now call Mobitz type II AV block.
Hay J. Bradycardia and cardiac arrhythmias produced by
depression of certain functions of the heart. Lancet
1906;1:138-143.
1906
Einthoven publishes the first organised presentation of normal
and abnormal electrocardiograms recorded with a string
galvanometer. Left and right ventricular hypertrophy, left and
right atrial hypertrophy, the U wave (for the first time),
notching of the QRS, ventricular premature beats, ventricular
bigeminy, atrial flutter and complete heart block are all
described. Einthoven W. Le telecardiogramme. Arch Int de Physiol
1906;4:132-164 (translated into English. Am Heart J
1957;53:602-615)
1906
Cremer records the first oesophageal electrocardiogram which he
achieved with the help of a professional sword swallower.
Oesophageal electrocardiography later developed in the 1970s to
help differentiate atrial arrhythmias. He also records the first
fetal electrocardiogram from the abdominal surface of a pregnant
woman. Cremer. Ueber die direkte Ableitung der Aktionstr?des
menslichen Herzens vom Oesophagus und ?as Elektrokardiogramm des
F?. Munch. Med. Wochenschr. 1906;53:811
1907
Arthur Cushny, professor of pharmacology at University College
London, publishes the first case report of atrial fibrillation.
His patient was 3 days post-op following surgery on an "ovarian
fibroid" when she developed a "very irregular" pulse at a rate
of 120 - 160 bpm. Her pulse was recorded with a "Jacques
sphygmochronograph" which shows the radial pulse pressure
against time - much like the arterial line blood pressure
recordings used in Intensive Care today. Cushny AR, Edmunds CW.
Paroxysmal irregularity of the heart and auricular fibrillation.
Am J Med Sci 1907;133:66-77.
1908
Edward Schafer of the University of Edinburgh is the first to
buy a string galvanometer for clinical use.
1909
Thomas Lewis of University College Hospital, London buys a
string galvanometer and so does Alfred Cohn of Mt Sinae
Hospital, New York.
1909
Nicolai and Simmons report on the changes to the
electrocardiogram during angina pectoris. Nicolai DF, Simons A.
(1909) Zur klinik des elektrokardiogramms. Med Kiln 5;160
1910
Walter James, Columbia University and Horatio Williams, Cornell
University Medical College, New York publish the first American
review of electrocardiography. It describes ventricular
hypertrophy, atrial and ventricular ectopics, atrial
fibrillation and ventricular fibrillation. The recordings were
sent from the wards to the electrocardiogram room by a system of
cables. There is a great picture of a patient having an
electrocardiogram recorded with the caption "The electrodes in
use".James WB, Williams HB. The electrocardiogram in clinical
medicine. Am J Med Sci 1910;140:408-421, 644-669
1911
Thomas Lewis publishes a classic textbook. The mechanism of the
heart beat. London: Shaw & Sons and dedicates it to Willem
Einthoven.
1912
Thomas Lewis publishes a paper in the BMJ detailing his careful
clinical and electrocardiographic observations of atrial
fibrillation. Lewis describes how he and a colleague, Dr
Woordruff a vet, identified the condition in horses and, at a
later date, witnessed the fibrillating heart of a horse on
Bulford Plain. "The chest was opened while the heart was still
beating, and I obtained, as did those with me, a clear view of a
fibrillating auricle, brought to this state, not by experimental
interference, but by disease." Lewis T. A Lecture ON THE
EVIDENCES OF AURICULAR FIBRILLATION, TREATED HISTORICALLY:
Delivered at University College Hospital. Br Med J 1912;1:57-60.
1912
Einthoven addresses the Chelsea Clinical Society in London and
describes an equilateral triangle formed by his standard leads
I, II and III later called 'Einthoven's triangle'. This is the
first reference in an English article I have seen to the
abbreviation 'EKG'.Einthoven W. The different forms of the human
electrocardiogram and their signification. Lancet
1912(1):853-861
1918
Bousfield describes the spontaneous changes in the
electrocardiogram during angina. Bousfield G. Angina pectoris:
changes in electrocardiogram during paroxysm. Lancet 1918;2:475
1920
Hubert Mann of the Cardiographic Laboratory, Mount Sinai
Hospital, describes the derivation of a 'monocardiogram' later
to be called 'vectorcardiogram'. Mann H. A method of analyzing
the electrocardiogram. Arch Int Med 1920;25:283-294
1920
Harold Pardee, New York, publishes the first electrocardiogram
of an acute myocardial infarction in a human and describes the T
wave as being tall and "starts from a point well up on the
descent of the R wave". Pardee HEB. An electrocardiographic sign
of coronary artery obstruction. Arch Int Med 1920;26:244-257
1924
Willem Einthoven wins the Nobel prize for inventing the
electrocardiograph.
1924
Woldemar Mobitz publishes his classification of heart blocks
(Mobitz type I and type II) based on the electrocardiogram and
jugular venous pulse waveform findings in patients with second
degree heart block. Mobitz W. Uber die unvollstandige Storung
der Erregungsuberleitung zwischen Vorhof und Kammer des
menschlichen Herzens. (Concerning partial block of conduction
between the atria and ventricles of the human heart). Z Ges Exp
Med 1924;41:180-237.
1926
A doctor from the Crown Street Women's Hospital in Sydney, who
wished to remain anonymous, resuscitates a new-born baby with an
electrical device later called a 'pacemaker'. The doctor wanted
to remain anoymous because of the controversy surrounding
research that artificially extended human life.
1928
Ernstine and Levine report the use of vacuum-tubes to amplify
the electrocardiogram instead of the mechanical amplification of
the string galvanometer. Ernstine AC, Levine SA. A comparison of
records taken with the Einthoven string galvanomter and the
amplifier-type electrocardiograph. Am Heart J 1928;4:725-731
1928
Frank Sanborn's company (founded 1917 and acquired by
Hewlett-Packard in 1961 and since 1999, Philips Medical Systems)
converts their table model electrocardiogram machine into their
first portable version weighing 50 pounds and powered by a
6-volt automobile battery.
1929
Sydney doctor Mark Lidwill, physician, and Edgar Booth,
physicist, report the electrical resuscitation of the heart to a
meeting in Sydney. Their portable device uses an electrode on
the skin and a transthoracic catheter. Edgar Booth's design
could deliver a variable voltage and rate and was employed to
deliver 16 volts to the ventricles of a stillborn infant.
Lidwell M C, "Cardiac Disease in Relation to Anaesthesia" in
Transactions of the Third Session, Australasian Medical
Congress, Sydney, Australia, Sept. 2-7 1929, p 160.
1930
Wolff, Parkinson and White report an electrocardiographic
syndrome of short PR interval, wide QRS and paroxysmal
tachycardias. Wolff L, Parkinson J, White PD. Bundle branch
block with short P-R interval in healthy young people prone to
paroxysmal tachycardia. Am Heart J 1930;5:685. Later, when other
published case reports were examined for evidence of
pre-excitation, examples of 'Wolff Parkinson White' syndrome
were identified which had not been recognised as a clinical
entity at the time. The earliest example was published by
Hoffmann in 1909. Von Knorre GH. The earliest published
electrocardiogram showing ventricular preexcitation. Pacing Clin
Electrophysiol. 2005 Mar;28(3):228-30
1930
Sanders first describes infarction of the right ventricle.
Sanders, A.O. Coronary thrombosis with complete heart block and
relative ventricular tachycardia: a case report, American Heart
Journal 1930;6:820-823.
1931
Charles Wolferth and Francis Wood describe the use of exercise
to provoke attacks of angina pectoris. They investigated the ECG
changes in normal subjects and those with angina but dismissed
the technique as too dangerous "to induce anginal attacks
indiscriminately". Wood FC, Wolferth CC, Livezey MM. Angina
pectoris. Archives Internal Medicine 1931;47:339
1931
first patented pacemaker Dr Albert Hyman patents the first
'artificial cardiac pacemaker' which stimulates the heart by
using a transthoracic needle. His aim was to produce a device
that was small enough to fit in a doctor's bag and stimulate the
right atrial area of the heart with a suitably insulated needle.
His experiments were on animals. His original machine was
powered by a crankshaft (it was later prototyped by a German
company but was never successful). "By March 1, 1932 the
artificial pacemaker had been used about 43 times, with a
successful outcome in 14 cases." It was not until 1942 that a
report of its successful short term use in Stokes-Adams attacks
was presented. Hyman AS. Resuscitation of the stopped heart by
intracardial therapy. Arch Intern Med. 1932;50:283
1932
Goldhammer and Scherf propose the use of the electrocardiogram
after moderate exercise as an aid to the diagnosis of coronary
insufficiency. Goldhammer S, Scherf D. Elektrokardiographische
untersuchungen bei kranken mit angina pectoris. Z Klin Med
1932;122:134
1932
Charles Wolferth and Francis Wood describe the clinical use of
chest leads. Wolferth CC, Wood FC. The electrocardiographic
diagnosis of coronary occlusion by the use of chest leads. Am J
Med Sci 1932;183:30-35
1934
By joining the wires from the right arm, left arm and left foot
with 5000 Ohm resistors Frank Wilson defines an 'indifferent
electrode' later called the 'Wilson Central Terminal'. The
combined lead acts as an earth and is attached to the negative
terminal of the ECG. An electrode attached to the positive
terminal then becomes 'unipolar' and can be placed anywhere on
the body. Wilson defines the unipolar limb leads VR, VL and VF
where 'V' stands for voltage (the voltage seen at the site of
the unipolar electrode). Wilson NF, Johnston FE, Macleod AG,
Barker PS. Electrocardiograms that represent the potential
variations of a single electrode. Am Heart J. 1934;9:447-458.
1935
McGinn and White describe the changes to the electrocardiogram
during acute pulmonary embolism including the S1 Q3 T3 pattern.
McGinn S, White PD. Acute cor pulmonale resulting from pulmonary
embolism: its clinical recognition. JAMA 1935;114:1473.
1938
American Heart Association and the Cardiac Society of Great
Britain define the standard positions, and wiring, of the chest
leads V1 - V6. The 'V' stands for voltage. Barnes AR, Pardee
HEB, White PD. et al. Standardization of precordial leads. Am
Heart J 1938;15:235-239
1938
Tomaszewski notes changes to the electrocardiogram in a man who
died of hypothermia. Tomaszewski W. Changements
electrocardiographiques observes chez un homme mort de froid.
Arch Mal Coeur 1938;31:525.
top
1939
Langendorf reports a case of atrial infarction discovered at
autopsy which, in retrospect, could have been diagnosed by
changes on the ECG. Langendorf R. Elektrokardiogramm bei
Vorhof-Infarkt. Acta Med Scand. 1939;100:136.
1942
Emanuel Goldberger increases the voltage of Wilson's unipolar
leads by 50% and creates the augmented limb leads aVR, aVL and
aVF. When added to Einthoven's three limb leads and the six
chest leads we arrive at the 12-lead electrocardiogram that is
used today.
1942
Arthur Master, standardises the two step exercise test (now
known as the Master two-step) for cardiac function. Master AM,
Friedman R, Dack S. The electrocardiogram after standard
exercise as a functional test of the heart. Am Heart J.
1942;24:777
1944
Young and Koenig report deviation of the P-R segment in a series
of patients with atrial infarction. Young EW, Koenig BS.
Auricular infarction. Am Heart J. 1944;28:287.
1947
Gouaux and Ashman describe an observation that helps
differentiate aberrant conduction from ventricular tachycardia.
The 'Ashman phenomenon' occurs when a stimulus falls during the
relative or absolute refractory period of the ventricles and the
aberrancy is more pronounced. In atrial fibrillation with
aberrant conduction this is demonstrated when the broader
complexes are seen terminating a relatively short cycle that
follows a relatively long one. The QRS terminating the shorter
cycle is conducted 'more aberrantly' because it falls in the
refractory period. The aberrancy is usually of a RBBB pattern.
Gouaux JL, Ashman R. Auricular fibrillation with aberration
simulating ventricular paroxysmal tachycardia. Am Heart J
1947;34:366-73.
1947
Claude Beck, a pioneering cardiovascular surgeon in Cleveland,
successfully defibrillates a human heart during cardiac surgery.
The patient is a 14 year old boy - 6 other patients had failed
to respond to the defibrillator. His prototype defibrillator
followed experiments on defibrillation in animals performed by
Carl J. Wiggers, Professor of Physiology at the Western Reserve
University. Beck CS, Pritchard WH, Feil SA: Ventricular
fibrillation of long duration abolished by electric shock. JAMA
1947; 135: 985-989.
Wiggers CJ, Wegria R. Ventricular fibrillation due to single
localized induction in condenser shock supplied during the
vulnerable phase of ventricular systole. Am J Physiol
1939;128:500
1948
Rune Elmqvist, Swedish engineer who had trained as a doctor but
never practiced, introduces the first ink jet printer for the
transcription of analog physiological signals. He demonstrates
its use in the recording of ECGs at the First International
Congress of Cardiology in Paris in 1950. The machine (the
mingograph) was developed by him at the company that later
became Siemens. (Luderitz, 2002)
1949
modern 'Holter' Monitor Montana physician Norman Jeff Holter
develops a 75 pound backpack that can record the ECG of the
wearer and transmit the signal. His system, the Holter Monitor,
is later greatly reduced in size, combined with tape / digital
recording and used to record ambulatory ECGs. Holter NJ,
Generelli JA. Remote recording of physiologic data by radio.
Rocky Mountain Med J. 1949;747-751.
1949
Sokolow and Lyon propose diagnostic criteria for left
ventricular hypertrophy i.e. LVH is present if the sum of the
size of the S wave in V1 plus the R wave in V6 exceeds 35 mm.
Sokolow M, Lyon TP. The ventricular complex in left ventricular
hypertrophy as obtained by unipolar precordial and limb leads.
Am Heart J 1949;37:161
1950
John Hopps, a Canadian electrical engineer and researcher for
the National Research Council, together with two physicians
(Wilfred Bigelow, MD of the University of Toronto and his
trainee, John C. Callaghan, MD) show that a coordinated heart
muscle contraction can be stimulated by an electrical impulse
delivered to the sino-atrial node. The apparatus, the first
cardiac pacemaker, measures 30cm, runs on vacuum tubes and is
powered by household 60Hz electrical current. Bigelow WG,
Callaghan JC, Hopps JA. "General hypothermia for experimental
intracardiac surgery." Ann Surg 1950; 1132: 531-539.
1953
Osborn, whilst experimenting with hypothermic dogs, describes
the prominent J (junctional) wave which has often been known as
the "Osborn wave". He found the dogs were more likely to survive
if they had an infusion of bicarbonate and supposed the J wave
was due to an injury current caused by acidosis. Osborn JJ.
Experimental hypothermia: respiratory and blood pH changes in
relation to cardiac function. Am J Physiol 1953;175:389.
1955
Richard Langendorf publishes the "rule of bigeminy" whereby
ventricular bigeminy tends to perpetuate itself. Langendorf R,
Pick A, Winternitz M. Mechanisms of intermittent ventricular
bigeminy. I. Appearence of ectopic beats dependent upon the
length of the ventricular cycle, the "rule of bigeminy."
circulation 1955;11:442.
1956
Paul Zoll, a cardiologist, uses a more powerful defibrillator
and performs closed-chest defibrillation in a human. Zoll PM,
Linenthal AJ, Gibson P: Termination of Ventricular Fibrillation
in Man by Externally Applied Countershock . NEJM 1956; 254:
727-729
1957
long QT syndrome Anton Jervell and Fred Lange-Nielsen of Oslo
describe an autosomal recessive syndrome of long-QT interval,
deafness and sudden death later known as the
Jervell-Lange-Nielsen syndrome. Jervell A, Lange-Nielsen F.
Congenital deaf mutism, functional heart disease with
prolongation of the QT interval and sudden death. Am Heart J
1957;54:59.
1958
Professor Ake Senning, of Sweden, places the first implantable
cardiac pacemaker designed by Rune Elmqvist into a 43-year-old
patient with complete heart block and syncope (Arne Larsson).
1959
Myron Prinzmetal describes a variant form of angina in which the
ST segment is elevated rather than depressed. Prinzmetal M,
Kennamer R, Merliss R, Wada T, Bor N. Angina pectoris. I. A
variant form of angina pectoris. Am J Med 1959;27:374.
1960
Smirk and Palmer highlight the risk of sudden death from
ventricular fibrillation particularly when ventricular premature
beats occur at the same time as the T wave. The 'R on T'
phenomenon. Smirk FH, Palmer DG. A myocardial syndrome, with
particular reference to the occurrence of sudden death and of
premature systoles interrupting antecedent T waves. Am J Cardiol
1960;6:620.
1963
Italian paediatrician C. Romano and Irish paediatrician O. Conor
Ward (the following year) independently report an autosomal
dominant syndrome of long-QT interval later known as the
Romano-Ward syndrome. Romano C, Gemme G, Pongiglione R. Aritmie
cardiache rare dell'eta pediatrica. Clin Pediatr.
1963;45:656-83.
Ward OC. New familial cardiac syndrome in children. J Irish Med
Assoc. 1964;54:103-6
1963
Excercise ECG Robert Bruce and colleages describe their
multistage treadmill exercise test later known as the Bruce
Protocol. "You would never buy a used car without taking it out
for a drive and seeing how the engine performed while it was
running," Bruce says, "and the same is true for evaluating the
function of the heart." Bruce RA, Blackman JR, Jones JW, Srait
G. Exercise testing in adult normal subjects and cardiac
patients. Pediatrics 1963;32:742
Bruce RA, McDonough JR. Stress testing in screening for
cardiovascular disease. Bull. N.Y. Acad Med. 1969;45:1288
1963
Baule and McFee are the first to detect the magnetocardiogram
which is the electromagnetic field produced by the electrical
activity of the heart. It is a method that can detect the ECG
without the use of skin electrodes. Although potentially a
useful technique it has never gained clinical acceptance, partly
because of its greater expense. Baule GM, McFee R. Detection of
the magnetic field of the heart. Am Heart J. 1963;66:95-96.
1966
Mason and Likar modify the 12-lead ECG system for use during
exercise testing. The right arm electrode is placed at a point
in the infraclavicular fossa medial to the border of the deltoid
muscle, 2 cm below the lower border of the clavicle. The left
arm electrode is placed similarly on the left side. The left leg
electrode is placed at the left iliac crest. Although this
system reduces the variability in the ECG recording during
exercise it is not exactly equivalent to the standard lead
positions. The Mason-Likar lead system tends to distort the ECG
with a rightward QRS axis shift, a reduction in R wave amplitude
in lead I and aVL, and a significant increase in R wave
amplitude in leads II, III and aVF. Eur Heart J. 1987
Jul;8(7):725-33
1966
Torsade de pointes FranºYs Dessertenne of Paris publishes the
first case of 'Torsade de pointes' Ventricular Tachycardia.
Dessertenne F. La tachycardie ventriculaire a deux foyers
opposes variables. Arch des Mal du Coeur 1966; 59:263
1968
Journal of Electrocardiography, the Official Journal of the
International Society for Computerized Electrocardiology and the
International Society of Electrocardiology, is founded by Zao
and Lepeschkin.
1968
Henry Marriott introduces the Modified Chest Lead 1 (MCL1) for
monitoring patients in Coronary Care.
1969
Rosenbaum reviews the classification of ventricular premature
beats and adds a benign form that arises from the right
ventricle and is not associated with heart disease. This becomes
known as the 'Rosenbaum ventricular extrasystole'. Rosenbaum MB.
Classification of ventricular extrasystoles according to form. J
Electrocardiol 1969;2:289.
1974
Jay Cohn, of University of Minnesota Medical School, describes
the 'syndrome of right ventricular dysfunction in the setting of
acute inferior wall myocardial infarction'. Cohn JN, Guiha NH,
Broder MI. Right ventricular infarction. Am J Cardiol
1974:33:209-214
1974
Gozensky and Thorne introduce the term 'Rabbit ears' to
electrocardiography. Rabbit ears describe the appearence of the
QRS complex in lead V1 with an rSR' pattern (good rabbit) being
typical of Right Bundle Branch Block and an RSr' (bad rabbit)
suggesting a ventricular origin i.e. ventricular ectopy /
tachycardia. Gozensky C, Thorne D. Rabbit ears: an aid in
distinguishing ventricular ectopy from aberration. Heart Lung
1974;3:634.
1976
Erhardt and colleagues describe the use of a right-sided
precordial lead in the diagnosis of right ventricular infarction
which had previously been thought to be electrocardiographically
silent. Erhardt LR, Sjogrn A, Wahlberg I. Single right-sided
precordial lead in the diagnosis of right ventricular
involvement in inferior myocardial infarction. Am Heart J
1976;91:571-6
1978
Dr Mieczyslaw (Michael) Mirowski and others file a US Patent
"Circuit for monitoring a heart and for effecting cardioversion
of a needy heart" (#4184493) which employs a transistor circuit
that analyses the ECG signal using a probability density
function. This allows an implantable defibrillator to detect
when heart rhythm changes from normal (with steep QRS slopes) to
abnormal ventriclar fibrillation. This development of
machine-interpretation of the ECG is essential for the safe
deployment of an automated defibrillator system and is reported
in Circulation. Mirowski M, Mower MM, Langer A, Heilman MS,
Schreibman J. A chronically implanted system for automatic
defibrillation in active conscious dogs. Experimental model for
treatment of sudden death from ventricular fibrillation.
Circulation 1978;58:90-94.
1988
Professor John Pope Boineau of Washington University School of
Medicine publishes a 30-year percpective on the modern history
of electrocardiography. Boineau JP. Electrocardiology: A 30-year
Perspective. Ah Serendipity, My Fulsome Friend. Journal of
Electrocardiology 21. Suppl (1988): S1-9
1992
Brugada syndrome Pedro Brugada and Josep brugada of Barcelona
publish a series of 8 cases of sudden death, Right Bundle Branch
Block pattern and ST elevation in V1 - V3 in apparently healthy
individuals. This 'Brugada Syndrome' may account for 4-12% of
unexpected sudden deaths and is the commonest cause of sudden
cardiac death in individuals aged under 50 years in South Asia.
The technology of the electrocardiogam, which is over 100 years
old, can still be used to discover new clinical entities in
cardiology. Brugada P, Brugada J. Right Bundle Branch Block,
Persistent ST Segment Elevation and Sudden Cardiac Death: A
Distinct Clinical and Electrocardiographic Syndrome. J Am Coll
Cardiol 1992;20:1391-6
1992
Cohen and He describe a new non-invasive approach to accurately
map cardiac electrical activity by using the surface Laplacian
map of the body surface electrical potentials. He B, Cohen RJ.
Body surface Laplacian ECG mapping. IEEE Trans Biomed Eng
1992;39(11):1179-91
1993
Mac 5000, 15-lead ECG Robert Zalenski, Professor of Emergency
Medicine, Wayne State University Detroit, and colleagues publish
an influential article on the clinical use of the 15-lead ECG
which routinely uses V4R, V8 and V9 in the diagnosis of acute
coronary syndromes. Like the addition of the 6 standardised
unipolar chest leads in 1938 these additional leads increase the
sensitivity of the electrocardiogram in detecting myocardial
infarction. Zalenski RJ, Cook D, Rydman R. Assessing the
diagnostic value of an ECG containing leads V4R, V8, and V9: The
15-lead ECG. Ann Emerg Med 1993;22:786-793
1999
Researchers from Texas show that 12-lead ECGs transmitted via
wireless technology to hand-held computers is feasible and can
be interpreted reliably by cardiologists. Pettis KS, Savona MR,
Leibrandt PN et al. Evaluation of the efficacy of hand-held
computer screens for cardiologists' interpretations of 12-lead
electrocardiograms. Am Heart J. 1999 Oct;138(4 Pt 1):765-70
2000
Physicians from the Mayo Clinic describe a new hereditary form
of Short QT syndrome associated with syncope and sudden death
that they discovered in 1999. Several genes have since been
implicated. Gussak I, Brugada P, Brugada J, et al. Idiopathic
short QT interval: a new clinical syndrome? Cardiology.
2000;94(2):99-102
2005
Danish cardiologists report the successful reduction in the time
between onset of chest pain and primary angioplasty when the ECG
of patients is transmitted wirelessly from ambulance to the
cardiologist's handheld PDA (Personal Digital Assistant). The
clinician can make an immediate decision to redirect patients to
the catheter lab saving time in transfers between hospital
departments. Clemmensen P, Sejersten M, Sillesen M et al.
Diversion of ST-elevation myocardial infarction patients for
primary angioplasty based on wireless prehospital 12-lead
electrocardiographic transmission directly to the cardiologist's
handheld computer: a progress report. J Electrocardiol. 2005
Oct;38(4 Suppl):194-8 |
| Article Source: |
| http://www.ecglibrary.com/ecghist.html |
|
Back To Top |
| ¡¡ |
|
The Art of The ECG / EKG Machine |
Electrocardiogram machines, or ECG and EKG machines, have
become some of the most common equipment used to record and
measure normal or abnormal action within the heart. The use of
EKG machines has become so common due to the wide variety of
heart disorders that an ECG machine will detect, along with the
ease and speed in which the test is accomplished. Some of the
disorders which can be detected by the ECG machine include:
valve disease, electrolyte disturbances, prior heart attacks,
coronary artery disease, thickening of the heart muscle, and a
broad range other possible cardiac problems.
An ECG machine works by recording the electrical action of the
muscles in the heart on a graph. Electrodes, usually around 10
of them, are attached to the person on various parts of the body
including the arms, legs, and chest. Some machines may have
12-leads of electrodes that attach to the body. The electrodes
then sense the electrical action from the heart and send those
signals out to the EKG machine to be recorded on a graph.
Usually, an ECG machine can perform a test within 5 minutes or
less. Physicians are trained to read the graph produced by an
ECG machine, and consider the size and length of each individual
part of the print out.
In addition, FAA certified commercial pilots must have an
electrocardiogram done on a periodic basis with results
transmitted to the Federal Aviation Administration. These EKGs
must be done on an FAA approved interpretive ECG machine, a list
of approved machines can be found here
New and reconditioned medical equipment can be found at
Absolutemed.com. Top of the line EKG machines can be found by a
wide variety of manufactures that include well known brands such
as GE, Marquette, HP, ACS, Burdick, Medgraphics, Ivy, and more. |
| Article Source: |
| http://www.absolutemed.com/Medical-Equipment/EKG-ECG-Machines?range=46%2C79%2C79 |
|
Back To Top |
| ¡¡ |
|
Electrocardiogram |
This is the most common test for heart conditions. It is a
simple painless test that takes about 10 minutes. An
electrocardiogram machine records your heart's rhythm onto paper
through sticky electrodes which are placed on your chest, arms
and legs. The recording will show if the heart muscle is damaged
or short of oxygen.
Exercise ECG (also called a treadmill test or exercise stress
test)
Some heart problems only appear when your heart needs to work
harder. You may need an Exercise ECG (a continuous ECG) to show
how your heart is coping when you are exercising, instead of
resting. While you walk on a treadmill, which will slowly get
faster and tilt up hill, your heart rhythm and blood pressure
will be recorded. This test takes about 10 minutes.
Holter monitoring (also called ambulatory ECG)
The Holter monitor is used to identify any heart rhythm
problems. It is a small, portable, battery powered ECG machine
worn at home over a 24-48 hour period. It will record your heart
rate and rhythm over this time and you will be asked to keep a
diary of what you do and any symptoms that you experience while
you are wearing the Holter monitor. At the end of the time
period, the monitor needs to be returned to the hospital or
clinic so the recorded information can be studied.
Echocardiogram
This test uses ultra sound (sound waves) to study the structure
of your heart and how the heart and valves are working. A probe
is passed over your chest and heart which sends out and records
these sound waves showing a moving image of your heart on a
computer.
Blood tests
Your doctor may arrange for you to have various blood tests. The
results will be used to help them understand how to treat your
heart disease.
Echocardiogram Stress Test
Stress echo is performed to see how your heart works while you
exercise. An echocardiogram taken while you rest, you then
exercise, and then another echo is done while your heart is
beating fast or if you are unable to exercise, you are given
medication (Dobutamine) via an intravenous (IV) needle in your
arm which makes your heart react as if your body was exercising.
Transoesphageal Echocardiogram (TOE)
A TOE is a special type of echocardiogram. Pictures of your
heart are taken by inserting a probe into your throat (oesphagus),
these pictures are clearer because the oesphagus is close to
your heart and there is no chest wall in the way.
The back of your throat will be sprayed with some numbing
solution and although you will be awake during the procedure,
you will be given some medication to make you feel relaxed and
sleepy. You will be able to eat and drink about two hours after
the test, your nurse will tell you when you can do this.
Occasionally people have a sore throat for 24 hours afterwards.
Cardiac Catheterisation (Angiography)
A small tube (catheter) is inserted into an artery, usually in
your groin or arm, and is threaded through to the part of the
aorta near your heart, where the coronary arteries start. A
special dye is injected through the catheter, into your
bloodstream. Using the dye as a highlight, x-ray pictures of the
heart and coronary arteries are taken. Click here to order your
FREE copy of A Guide to Coronary Angiography.
Electrophysiological studies (EPS)
If you have abnormal heart rhythms (arrythmias) or palpitations
you may need this test. Similar to an angiography, fine tubes
(electrode catheters) are fed into a vein and/or artery usually
in the groin. They are then gently moved into the heart, where
they stimulate the heart and record your heart's electrical
activity.
Tilt Table Test
If you have episodes of fainting, a tilt table test is used to
investigate if these could be related to your heart. You lie on
a special table, which can be angled so you lie down or stand up
and you will be attached to a heart and blood pressure monitor
which record how your heart rate and blood pressure respond to
changes in position. During the test you may have an intravenous
needle in your arm so you can be given medication.
CT Angiography
A CT machine is used to produce a detailed picture of your
heart. A needle is put into your arm so dye can be injected into
your blood system. The dye briefly fills the arteries and heart
so they can be seen on x-ray. The pictures give your doctor
information about how your heart and blood vessels are working
and show up any areas which are blocked or narrowed. |
| Article Source: |
| http://www.heartfoundation.org.nz/index.asp?pageID=2145859272 |
|
Back To Top |
| ¡¡ |
|
Interpreting Electrocardio-graph |
An ECG is printed on paper covered with a grid of squares.
Notice that five small squares on the paper form a larger
square. The width of a single small square on ECG paper
represents 0.04 seconds. To successfully interpret ECGs, you
must have this value committed to memory. Do this now. If each
small square represents 0.04 seconds, then a second will be 25
small squares across. If you print out a minute's worth of your
heart's electrical activity, the paper would be 1500 small
squares wide. If something on an ECG is, let's say, 12 small
squares in width, that means that it lasted 12 x 0.04, or almost
half a second. A common length of an ECG printout is 6 seconds;
this is known as a "six second strip."
Each one of the figures represents an ECG pattern displaying
three types of abnormal rhythms: Tachycardia, Bradycardia, and
Arrhymthmia. Identify each.


 |
| Article Source: |
| http://www.biologycorner.com/anatomy/circulatory/ecg.html |
|
Back To Top |
| ¡¡ |
|
Electrocardio-graph description |
General Description
As the heart undergoes depolarization and repolarization, the
electrical currents that are generated spread not only within
the heart, but also throughout the body. This electrical
activity generated by the heart can be measured by an array of
electrodes placed on the body surface. The recorded tracing is
called an electrocardiogram (ECG, or EKG). A "typical" ECG
tracing is shown to the right. The different waves that comprise
the ECG represent the sequence of depolarization and
repolarization of the atria and ventricles. The ECG is recorded
at a speed of 25 mm/sec, and the voltages are calibrated so that
1 mV = 10 mm in the vertical direction. Therefore, each small
1-mm square represents 0.04 sec (40 msec) in time and 0.1 mV in
voltage. Because the recording speed is standardized, one can
calculate the heart rate from the intervals between different
waves.
P wave
The P wave represents the wave of depolarization that spreads
from the SA node throughout the atria, and is usually 0.08 to
0.1 seconds (80-100 ms) in duration. The brief isoelectric (zero
voltage) period after the P wave represents the time in which
the impulse is traveling within the AV node (where the
conduction velocity is greatly retarded) and the bundle of His.
Atrial rate can be calculated by determining the time interval
between P waves. Click here to see how atrial rate is
calculated.
The period of time from the onset of the P wave to the beginning
of the QRS complex is termed the P-R interval, which normally
ranges from 0.12 to 0.20 seconds in duration. This interval
represents the time between the onset of atrial depolarization
and the onset of ventricular depolarization. If the P-R interval
is >0.2 sec, there is an AV conduction block, which is also
termed a first-degree heart block if the impulse is still able
to be conducted into the ventricles.
QRS complex
The QRS complex represents ventricular depolarization.
Ventricular rate can be calculated by determining the time
interval between QRS complexes. Click here to see how
ventricular rate is calculated.
The duration of the QRS complex is normally 0.06 to 0.1 seconds.
This relatively short duration indicates that ventricular
depolarization normally occurs very rapidly. If the QRS complex
is prolonged (> 0.1 sec), conduction is impaired within the
ventricles. This can occur with bundle branch blocks or whenever
a ventricular foci (abnormal pacemaker site) becomes the
pacemaker driving the ventricle. Such an ectopic foci nearly
always results in impulses being conducted over slower pathways
within the heart, thereby increasing the time for depolarization
and the duration of the QRS complex.
The shape of the QRS complex in the above figure is idealized.
In fact, the shape changes depending on which recording
electrodes are being used. The shape will also change when there
is abnormal conduction of electrical impulses within the
ventricles. The figure to the right summarizes the nomenclature
used to define the different components of the QRS complex.
ST segment
The isoelectric period (ST segment) following the QRS is the
time at which the entire ventricle is depolarized and roughly
corresponds to the plateau phase of the ventricular action
potential. The ST segment is important in the diagnosis of
ventricular ischemia or hypoxia because under those conditions,
the ST segment can become either depressed or elevated.
T wave
The T wave represents ventricular repolarization and is longer
in duration than depolarization (i.e., conduction of the
repolarization wave is slower than the wave of depolarization).
Sometimes a small positive U wave may be seen following the T
wave (not shown in figure at top of page). This wave represents
the last remnants of ventricular repolarization. Inverted or
prominent U waves indicates underlying pathology or conditions
affecting repolarization.
Q-T interval
The Q-T interval represents the time for both ventricular
depolarization and repolarization to occur, and therefore
roughly estimates the duration of an average ventricular action
potential. This interval can range from 0.2 to 0.4 seconds
depending upon heart rate. At high heart rates, ventricular
action potentials shorten in duration, which decreases the Q-T
interval. Because prolonged Q-T intervals can be diagnostic for
susceptibility to certain types of tachyarrhythmias, it is
important to determine if a given Q-T interval is excessively
long. In practice, the Q-T interval is expressed as a "corrected
Q-T (QTc)" by taking the Q-T interval and dividing it by the
square root of the R-R interval (interval between ventricular
depolarizations). This allows an assessment of the Q-T interval
that is independent of heart rate. Normal corrected Q-Tc
intervals are less than 0.44 seconds.
There is no distinctly visible wave representing atrial
repolarization in the ECG because it occurs during ventricular
depolarization. Because the wave of atrial repolarization is
relatively small in amplitude (i.e., has low voltage), it is
masked by the much larger ventricular-generated QRS complex.
ECG tracings recorded simultaneous from different electrodes
placed on the body produce different characteristic waveforms. |
| Article Source: |
| http://www.cvphysiology.com/Arrhythmias/A009.htm |
|
Back To Top |
| ¡¡ |
|
The electrocardiogram |
Definition
An electrocardiogram (ECG) is a test that records the electrical
activity of the heart.
See also:
Holter monitoring.
Stress test
Why is the Test Performed?
An ECG is used to measure:
Any damage to the heart
How fast your heart is beating and whether it is beating
normally
The effects of drugs or devices used to control the heart (such
as a pacemaker)
The size and position of your heart chambers
An ECG is a very useful tool for determining whether a person
has heart disease. Your doctor may order this test if you have
chest pain or palpitations.
An ECG may be included as part of a routine examination in
patients over age 40.
How is the Test Performed?
You will be asked to lie down. The health care provider will
clean several areas on your arms, legs, and chest, and then
attach small patches called electrodes to the areas. It may be
necessary to shave or clip some hair so the electrodes stick to
the skin.
The number of patches used may vary.
You usually need to remain still, and you may be asked to hold
your breath for short periods during the procedure. It is
important to be relaxed and relatively warm during ECG
recording. Any movement, including muscle tremors such as
shivering, can alter the results.
The electrodes are connected by wires to a machine that converts
the electrical signals from the heart into wavy lines, which are
printed on paper and reviewed by the doctor.
Sometimes this test is done while you are exercising or under
minimal stress to monitor changes in the heart. This type of ECG
is often called a stress test.
How to Prepare for the Test?
Make sure your health care provider knows about all the
medications you are taking, as some can interfere with test
results.
Exercising or drinking cold water immediately before an ECG may
cause false results.
How will the Test Feel?
An ECG is painless. No electricity is sent through the body. The
electrodes may feel cold when first applied. In rare cases, some
people may develop a rash or irritation where the patches were
placed.
Risks
There are no risks. No electricity is sent through the body, so
there is no risk of shock.
Special Considerations
The accuracy of the ECG depends on the condition being tested.
Some heart conditions are not detectable all the time, and
others may never produce any specific ECG changes.
A person who has had a heart attack or who may have heart
disease may need more than one ECG. There is no reason for
healthy people to have yearly routine testing unless they have a
family or person history of specific heart diseases or other
medical conditions.
Normal Values
▪Heart rate: 50 to 100
beats per minute
▪Heart rhythm: consistent
and even
▪What do Abnormal Results
Mean?
▪Abnormal ECG results may
be a sign of
Abnormal heart rhythms (arrhythmias)
▪Cardiac muscle defect
▪Congenital heart defect
▪Coronary artery disease
▪Ectopic heartbeat
▪Enlargement of the heart
▪Faster-than-normal heart
rate (tachycardia)
▪Heart valve disease
▪Inflammation of the heart
(myocarditis)
▪Changes in the amount of
electrolytes (chemicals in the blood)
▪Past heart attack
▪Present or impending
heart attack
▪Slower-than-normal heart
rate (bradycardia) |
| Article Source: |
| http://health.allrefer.com/health/ecg-ecg.html |
|
Back To Top |
| ¡¡ |
|
History and Term Electrocardiogram |
In 1787 Galvani was the first to discover the relationship
between electrical currents and muscle contractions. In 1843
Carlo Matteucci detected that the heart¡¯s activity is also based
on electrical currents. The first graphic representation of this
was made by E.J. Marey in 1876. The breakthrough came with the
Dutch physiologist Willem Einthoven who was awarded the Nobel
Prize in medicine for the invention of the electrocardiography.
The deflections and curve descriptions developed by him are
still in use today. These deflections were amended by the
American cardiologist Emanuel Goldberger on limb leads and by
Frank Wilson on precordial leads.
The electrocardiogram records the electrical activity of the
heart. The heart is a muscular organ which beats in rhythm to
pump the blood through the body. The signals that make the
heart¡¯s muscle fibres contract come from the sinoatrial node,
which is the natural pacemaker of the heart.
In an ECG test, the electrical impulses made while the heart is
beating, are recorded and usually shown on a piece of paper.
This is known as an electrocardiogram, and provides information
on the condition and performance of the heart.
Anatomy of the Heart
It is important to know the heart¡¯s structure and blood flow to
understand the ECG. The heart is a hollow muscle which is
divided into four chambers. These are the right atrium, the
right ventricle and the left atrium and the left ventricle.
The right atrium receives venous blood which passes via the
tricuspid valve to the right ventricle, which propels it through
the pulmonary artery to the lungs. In the lungs venous blood
comes in contact with inhaled air, picks up oxygen, and loses
carbon dioxide. Oxygenated blood is returned to the left atrium
through the pulmonary veins. Passage of blood through the left
atrium, bicuspid valve, into the left ventricle. Via the aortic
valve the blood is pumped in the aorta and the arterial branches
of the whole body.
Standard ECG - Records
Method of graphic tracing of the electric current generated by
the heart and information on the condition and performance of
the heart.
In the following the waveform of a normal ECG is explained. Any
deviation from the norm in a particular electrocardiogram is
indicative of a possible heart disorder. A selection of ECG
recordings taken during various electrocardiographic tests will
be explained.
The normal ECG shows characteristic waves in its course. It was
Einthoven who assigned the letters P, Q, R, S, and T to the
various deflections.
The P-Wave
The P-Wave is caused by atrial contraction. The first upward
deflection corresponds with the right atrium and the second
downward deflection corresponds with the left atrium.
Examples of deviations from the normal P-Wave indicate:
Pointed, upright P-wave when the right atrium is overstrained,
e.g. in case of cor pulmonale acutum or cor pulmonale chronicum
, i.e. in Latin pulmonary heart ¨C a pressure-loaded heart due to
a risen pressure in the pulmonary circulation because of a
pulmonary disease
Bicuspidal , often spreadout P-wave, emphasizing the 2nd peak,
e.g. indicating high blood pressure
Both parts of the P-wave are changed, merged representation of
the changed P-wave as mentioned before, e.g. in case of high
blood pressure, right heart hypertrophy and severe organic heart
defects
Negative deflection of the P-wave occurs in cases of pacemaker
actions in the atrioventricular area
ECG-Instruments measure the PR-Interval
The P-Q-time or PR-Interval extends from the start of the P-wave
to the very start of the QRS-complex. The excitation is
decreased by the AV-node and led via the bundle of His to the
left and right bundle branch (thus, conduction time).
The normal duration is between 0.12 ¨C 0.20 sec. A PR-interval of
more than 0.20 sec may indicate a first degree an AV-block.
The QRS- Complex
The excitation is led via the left bundle branch and the
ventricular septum and is visible as Q-wave n the ECG. During
the R-phase most of the heart¡¯s muscles are activated. For this
reason the ECG shows the great wave.
Whereas during the S-phase the activation runs from the apex of
heart to the base of the right and left ventricle.
Examples for Abnormalities of the QRS-Complexes
Left-ventricular hypertrophy demonstrates thickening of the
heart muscle (left ventricle). More cells are activated which
leads to a taller R-wave.
Right-ventricular hypertrophy demonstrates thickening of the
heart muscle (right ventricle). However, the right ventricle
still has less muscle mass than the left ventricle.
Ventricular conduction abnormality. An abnormal QRS-complex can
be a sign of disturbances in stimulus conduction. There is also
an abnormal QRS-complex, which may indicate myocardial
infarction.
The ECG-instrument also measures the QRS-duration from the
beginning of the Q-wave to the end of the S-wave.
It demonstrates the duration of the depolarization of the
heart¡¯s ventricles. A normal duration lies between 0.08 and 0.12
sec. If the duration is longer this may indicate a conduction
abnormality as described before.
ECG-instruments measure the QT/QTc Interval
The QT-interval is measured from the beginning of the Q-wave to
the end of the T-wave. The QT-interval represents the duration
of activation and recovery of the ventricular muscles. This
duration is reciprocal to the pulse.
The ST-Segment
The ST-segment represents the period from the end of ventricular
depolarization to the beginning of ventricular repolarization.
Here all cells of the atria are depolarized. An isoelectric line
is generated because in this segment there is no electrical
current.
Examples of Abnormalities of the ST-segment
A depressed ST-segment can be a sign of heart insufficiency
A low elevation reveals e.g. bradycardia
A clear St-segment elevation may indicate e.g. acute myocardial
infarction
ECG-Instruments measure the ST-Segment and may detect the
before-mentioned coronary diseases.
The T-Wave
The T-wave represents the repolarization of the ventricles and
runs into the same direction as the R-wave.
T-wave Abnormalities
A flattened T-wave indicates e.g. vegetative dystonia
Negative (or inverted) T-waves can be a sign of heart
insufficiency
Hyperacute T-wave is sometimes an early sign of myocardial
infarction in the form of a tall positive symmetrical T- wave
The U-Wave
The U-wave is typically small and follows the T-wave. Its origin
has not yet been understood completely. |
| Article Source: |
| http://www.davita-shop.co.uk/ecg-instruments.html |
|
Back To Top |
| ¡¡ |
|
|
|